Chapter 2 Morgagni 268

2.1. High Tension goes off and there is error message saying “Leakage! HT allowed after 1 minute”. What can I do?

2.1. High Tension goes off and there is error message saying “Leakage! HT allowed after 1 minute”. What can I do?

You can simply click on button Empty Buffer on Vacuum module. This will empty buffer tank and keep pressure in the tank low, which helps to maintain sufficient pressure difference between front and back of ODP pump. The temperature of ODP pump will increase after emptying the buffer tank.

Empty Buffer manually once a while, e.g. every 2 hours, will avoid the leakage problem.