About Us

Home

Sign Up

The Instructor

EM Lab Policies

Links

Rosenstiel

Keck Institute

Faculty & Lab

EM LAB Policies

 

   
 

INDEX

  1. Priority for scheduling time on the Phillips CM12, 420 and 301
  2. Computer sign-ups for the microscopes and supporting equipment
  3. Training
  4. GATAN 651N anti-contaminators and cryo-holders
  5. Plunger
  6. Negative developing darkrooms and the printing darkroom
  7. General Policies
   
 

1. Priority for scheduling time on the Philips CM12, 420 and 301 is as follows:

a. Each user is limited to 2 sign-ups (as indicated below) within any sign-up period.

b. Reservations for the CM12 and 420T: Spot-scanning takes priority on the CM12 followed by cryo-microscopy, low dose and high dose on both instruments. Time is reserved on four hour basis from 8am - 12am, 12am - 4pm and 4pm - 8pm. If no one is signed up to use the microscope after your session you can extend your time. If the microscopes are not being heavily used you may sign up for the entire day but this is the exception, not the rule.

c. Reservations for the 301: Each user is limited to two hour time slots per week (prime time from 8am - 5pm) during periods of heavy use only. If you are 15 minutes late, your reserved time is open to another user unless you have made previous arrangements. As always, the first user of the day is responsible for turning the instrument on and the last user of the day is responsible for turning it off.

2. Computer sign-ups for the microscopes and supporting equipment: a. Before you can signup to use any of the electron microscopes you must have submit a charge line to Dr.Chen Xu. Once your account has been established you will have access to the Keck web site, which contains instructions for the research initiatives as well as the calendars for instrument reservations.

b. If you know in advance that you will not be using your reserved time, please remove your name from the reservation calendar as soon as possible so that others may use that time.

3. Training in the use of the microscopes, supporting equipment and darkrooms should be scheduled with a staff member of the EM Lab.

4. GATAN 651N anti-contaminators and cryo-holders:

a. Pump the 651N anti-contaminators at 90° C overnight followed by pumping to room temperature before closing the vacuum valve and filling the dewar with liquid nitrogen for use. These dewars should be pumped once a week (or approximately every 6 days of use). You will know they need to be pumped when the outside of the feels cold and damp and the temperature is unstable at the minimum value of -180 °C. (Instructions for pumping the 651N anti-contaminators are located in the instruction manual for RI4).

b. For best results, pump the 626 cryo-holders for 3.5 to 4 hours at 135 °C and allow a cool down pumped time of 0.5 hours. Use the GATAN 655 Dry Pumping Station to pump the cryo-holders. (Instructions for pumping the GATAN 626 cryo-holders are taped to the top of the pumping station.)

c. Log all pertinent information regarding cryo-holder and anti-contaminator maintenance in the cryo log book.

5. Plunger: a. Reservations may be made for the Grigorieff plunger and the EM lab pneumatic plunger on the Keck web calendar. Any unused propane and ethane should be vented in the hood near the plunger.

b. After plunging your specimens, clean and return all equipment to its proper storage location.

c. If you use the humidifier for the EM lab plunger, please use only distilled water in the reservoir. Be sure to empty and dry the humidifier when finished.


6. Negative developing darkrooms and the printing darkroom:

a. Please be considerate and clean up after yourself.

b. Photographic paper is available in the stock room on the second floor.

c. Handle EM film with care to avoid scratches and static discharges. If you inadvertently damage a piece if film, discard it rather than loading it into the camera for the next user.

d. The 1st floor EM lab and darkroom should be locked at all times unless you are working in these areas. If you are the last one out for the day, make sure the rooms are locked.

7. General policies

a. VERY IMPORTANT: If you experience problems while operating any of the instruments, please record this information in the log book and notify Dr.Chen Xu. Attempting to correct a malfunction and inadvertently damaging one of the microscopes can void our service agreement with OpTechs, Inc. (301) or NanoTech Consulting (420T and CM12). In the event of an emergency, shut down procedures for the microscopes and ancillary instruments are located in the instruction manuals for each.

b. Information regarding routine and emergency service as well as other EM lab related activities will be posted on the bulletin board of the Keck web site. You should always expect last minute changes in the microscope service schedules.

c. In the event that the first floor air handling unit malfunctions, please follow the directions for shut down of the unit (they are posted on the back of the main door to the 1st floor EM suite) and contact Linda Melanson or Eddie Arocco for assistance. Very high temperature in the room (above 80° C) indicates a problem.

d. All users must clean up their work areas once finished and return all items used to their proper storage location as a courtesy to the next user.

e. All chemical waste must be disposed of in accordance with the rules of the Brandeis University Safety Office. If you have questions regarding this, please contact an EM lab staff member for assistance, Bob Elias (Safety Management) at extension. 64262 or Robin Bell (Radiation Safety) at extension 64261.

f. VERY IMPORTANT: Confine the use of heavy metal stains to the staining lab bench in room 436. Place any contaminated materials into the waste containers provided. All solid and liquid waste must be properly disposed of using the containers provided next to the sink. DO NOT leave any contaminated waste materials out on the counter (this includes partially used vials of uranyl acetate).

g. The CM12, 420, 301, Edwards Vacuum Evaporator, EmiTech Glow Discharge Station, GATAN CCD camera and GATAN Dry Pumping Station should be left in their respective stand-by settings for the next user. This information is posted in the microscope rooms and in the instruction manuals located near each instrument. If you have questions about this please ask for assistance.

h. Conserve Energy. Turn off the lights and equipment as the situation dictates.

 

 

Back to Top

 

Copyright © BrandeisEM. All rights reserved