"Shuttle
Device for Use in a Shared Commercial NMR Instrument Version II "
(December
2004)
by A. G. Redfield, Brandeis University
Return
to Index
Section
SH Shuttle, NMR tube, and plugs
The shuttle etc design were
changed from MRC to make the nmr tube and adapter shorter and the rest
simpler to make. The main changes are: the nmr tube is shorter and the
adapter is 1/2 in diameter instead of 3/8 (these things improve rigidity);
it attaches to the shuttle body externally in a simpler way; the O-rings
of the debounce system ride on the shuttle body; the shuttle is hollow
so that the entire shuttle plus adapter plus tube now weighs 30 gm instead
of 60. To allow the body to be hollow I dispensed with the extra assembly
inside the shuttle body that acts as an extra shock absorber and find
no tube breakage in several recent runs (8 mm). See fig
SH-1.
This design has now (12/04)
been used frequently with both 5 mm and 8 mm NMR tubes and I am confident
that it is an improvement over version I.
The shuttle body consists
of a simple tube, made as thin as is easy, about 3/4 OD and 5/8 ID drilled
with a carbide drill, ends reasonably parallel and Id at the ends turned
slightly larger to provide a fairly well aligned precise fit to be epoxied
to the 2 end caps. The end caps carry the two precise bearing surfaces
0.798 OD that align the shuttle inside the 0.800 glass shuttle tube,
and 1/2 inch extensions that hold the seven 1/2 x 3/4 inch Viton (fluorocarbon)
O rings for shock absorption, on each end. The upper cap is drilled
out to save weight and the bottom one (otherwise identical to the top
one) mates to the NMR tube adapter. You need at least 2 shuttle bodies,
the second one to use with the vee block while running if you want to
seal a new sample. I will return to the shuttle body and vee block later.
Sample loading See
fig. SH-1.
The NMR tube design is similar
to MRC but the tube is shorter, about 80 mm long. I plan to get 9 inch
tubes from now on and have learned to cut them in two. Well, perhaps.
(I will explain how later). I pipette about 0.1 ml of low-viscosity
optical grade epoxy (Epotek 405 from Epoxy Technology, Billerica MA)
into each 5 mm thin-wall tube a day or more before running, using a
broken-off Wilmad long pipette. (about 0.25 ml of epoxy in the 8 mm
tubes). I withdraw the pipette very rapidly to try to get as little
epoxy on the walls. I spin the tubes in a small clinical centrifuge
to get most of the epoxy off the walls and cure at room temperature
for at least 3-4 hours before putting in the sample. I expect to use
the upper half of each cut off tube to make a second tube by plugging
the upper half with an epoxy plug-- by sealing off the good upper end
with parawax and pouring epoxy as above in from the other end to form
a sealed bottom about 5 mm or a little more long. I hope this works,
the 8 mm tubes are expensive. (This is easy but I have not yet had the
courage to use thse home-made tubes). I then cut off the NMR tube to
the correct length son that it can later be epoxied into the adapter,
assuming that the nmr sample will be 15 mm long and that the center
of the sample will be 65 mm below the bottom end of the adapter and
that about 12 mm is needed for the epoxy coupling to the adapter. Thus
the NMR tube will be about 5+(15/2)+65+10=87.5 mm long, but this is
varied depending on how long the poured epoxy plug at the bottom is,
and it is longer for the 8 mm tubes because the bottom epoxy plug has
to be longer. The main point is to try to make the center of the sample
space (which is 15 mm long) be 65 mm below the lower end of the NMR
tube adapter (Fig. SH-1). I designed
the adapter (Fig. SH-5) so that there
is excess length for the NMR tube to poke into, and so that it is not
required to be very precise in cutting it off.
It is rather tedious, and
risky to the tube, to clean off the epoxy as just described, and I have
gone back to using a shield tube (see MRC) to avoid getting epoxy on
the upper walls.
Details: After the
NMR tube has been shortened to the correct length (above), break off
a long Wilmad NMR pipette so that its end reaches to the bottom of the
tube at least. Cut off (or re-use) a piece of shield tube that slides
over the end of the pipette, far enough to leave about 1 cm of the lower
end of the pipette exposed. This tube has to be small enough in diameter
to fit into the 5 mm tube. The only kind of tubing that I have found
suitable is Intramedic-Clay-Adams Beckton Dickinson number 42745 polyethylene
tube sold by several large chemistry supply houses, ID 3.17 mm, OD 3.99
mm. You can prepare at least two, probably four, tubes at a time using
EPOTEK 305 (Epoxy Technology, Billerica MA) resin which sets very slowly
(hours) and has low viscosity. Set up an ordinary 1 ml pipettor vertically,
with its standard plastic pipette tube attached in a standard lab stand
and clamp. Couple the glass pipette to the plastic pipette tip with
a short piece of rubber tube. You will soon develop your own variations,
but it seems best to have the end of the glass pipette about 3 cm above
the base. Mix about 1 ml of epoxy per two NMR tubes in a short dispo
test tube. The shield tube should now be slid up over the bottom of
the glass pipette, with 1-2 cm of the glass end uncovered at the bottom.
Suck 1 ml of epoxy into the glass tube in the usual way, but keep the
tube containing the epoxy low enough so that epoxy does not touch or
cover the shield tube. Release the top button of the pipettor slowly
while sucking in the epoxy, to avoid excess vacuum. Do not worry about
small bubbles that will appear in the epoxy. Now try to have a short
(mm's) length of air space at the bottom of the glass pipette, and wipe
off the pipette and the shield tube thoroughly to remove excess epoxy.
Now insert the glass pipette with shield into the NMR tube and push
the pipettor button slowly to release the desired amount of epoxy, releasing
close to the bottom of the epoxy but again trying not to have the epoxy
touch the shield. I usually put in epoxy to a depth equal to the nmr
tube diameter. Finally, grasp the top of the NMR tube in such a way
as to keep the shield tube fixed relative to it, and pull up the glass
tube so that its tip is just within the shield. Then shift to grasp
the shield and glass pipette and pull the NMR tube off of both. This
usually works; if not, little blobs of epoxy will be left on the upper
part of the NMR tube. (Use a wooden stick or a small NMR tube inverted
to try later to scrape them off so that the upper hollow susceptibility
plug will slide in without excess force.) Now do another tube by first
pushing the shield tube back up and wiping carefully.
I have previously prepared
the hollow upper susceptibility plug by scraping the epoxy off its upper
end, left over from the previous use, with a new single-edge razor blade,
and perhaps also filing or sanding it gently. I rely on the fact that
PEEK is much harder than epoxy so that it is only slightly worn by scraping;
in fact I have yet to throw away any of these plastic pieces because
they were too worn. I test it in the course of this cleaning to see
that it will fit in an old NMR tube of the same size and then test that
it will fit in the tube I plan to use. If not, because of epoxy left
on the inside of the new tube, I clean this off first with a Fisher
brand wooden stick and then with a smaller NMR tube stuck inside the
larger one, i.e a 4 mm, scraping inverted using the top as a chisel,
inside a 5 mm tube. And then wash out the epoxy pieces with water and
ethanol).
The piece of wire at the
upper end of the upper plug is now used as a spring to retard the plug
from falling to the bottom during a spin. I bend it sideways carefully
so as to do this but not break the tube, and test by putting it slowly
and inverted in the junk NMR tube part way upside down. About 20 gram
or more should be needed to move it. This is measured (not for every,
tube but occasionally) by putting the plug with wire into the top of
a spare NMR tube, and invert it with a wooden or glass rod stuck up
into the top. Then I press the rod on to an electronic top-loader balance
and push until the plug moves; the force it indicates when the plug
moves should be around 100 times the weight of the plug. I bend a little
hook on the end of this wire so that a curved surface slides along the
glass.
One adapter (new design,
fig. SH-5) has to be prepared for each
sample to be sealed. If newly made it has to be de-greased with ethanol.
If previously used the remains of the previous NMR tube is broken off
crudely and as much removed from the lower end of the adapter with miniature
pliers. Then I crudely break out the glass from the hole with the largest
size drill that almost fits this hole, holding the drill bit with pliers
and stopping when the drilling feels smoother, or using a good quality
electric drill to turn it at low speed. I repeat this with a larger
drill bit or two, and finally apply a 5 mm reamer. (I do not have an
8 mm reamer and use a drill in what follows for the 8 mm size.) I insert
the reamer or final drill and turn it while exerting lateral force to
scrape off the epoxy from the sides of the hole. I test by seeing if
the bottom end of an nmr tube slides easily into the hole and is not
pushed away from concentric by remaining pieces of epoxy in the hole.
It should be possible to wiggle it enough to have the end of the NMR
tube slide easily into the Vee-block adapter suitable for the NMR tube,
when the tube inserted into the adaptor and connected to the shuttle
are in the V-block flat on the bench.
The above steps take only
ten minutes or less per sample, after practice. Before the next step
you also have to locate the vee-block adapters and vee block (see below,
and fig SH 2) and mentally go over
what follows, which has to be done with moderate speed before the epoxy
solidifies. The vee block should be ready with rubber bands to hold
the shuttle assembly with the correct spacer held on in place with paper
tape. Buy a bag of wide rubber bands. You have to set up the gas bag
if needed (see MRC) and figure out how to use it (a pain). The upper
plug should be washed in distilled water and dried with ethanol and
in some cases soaked in buffer of some type suggested by the biochemistry
(i. e. containing EDTA and/or DTT) and finally dried and placed ready
to be loaded on a new kimwipe. You have to find a #16 syringe tip (Fisher,
BD 16G 1 1/2) and a 5 or 10 cc syringe body and assemble them tightly
ready on the bench, for transferring epoxy (a smaller syringe tip is
inconvenient though possible. Order a box of the #16). Cut off the very
sharp end of the syringe with small diagonal cutters for your safety.
Find and set up the epoxy ready to mix (below).
If bubbles are of great concern
I degas the sample as described in MRC in the gas bag, requiring about
1/2 hour using my slow shaker (MRC) under helium. Usually I do not do
so for phospholipid vesicles for which the line-width is 50 Hz or more.
I did try to degas epoxy but gave it up, though it could be done. See
MRC for more on this.
I then pipette in the sample
with a Wilmad long NMR pipette, in a gas bag if required, or after being
chelexed if needed (for 31P, if EDTA can't be
used). About 270 microliters for the 5 mm tubes, 750 for the 8 mm. It
is important not to put in too much sample. If you do, pipette enough
out so that the sample length is now about 17-8 mm. The extra length
(over the final 15 mm length) fills the space between the upper plug
and the inner wall of the NMR tube. After this the sample is centrifuged
("spun") for a few minutes to avoid excess moisture on the walls of
the glass especially at the top. A new pipe cleaner can also be used
to dry the tube. Then check the height of the sample and adjust the
volume as mentioned, if needed.
Now move to a window or other
strong light and put in the upper plug in the tube with the wire up.
It may be worth practicing what follows with an inexpensive colored
sample like cytochrome C. Shove the plug down carefully til it hits
the surface of the sample and then a very little further. There will
be a large bubble at the top of the sample. In this and what follows
the object is to get the top meniscus of the bubble-free aqueous sample
a few millimeters below the groove at the top of the plug. (see MRC,
the groove blocks the flow of aqueous solution and buffer across it
so they do not mix) and never into the groove or especially into
the space above the groove which is where the epoxy is supposed to go.
(If this happens I generally spin to tube at the highest speed possible
to get the aqueous sample back down). So look at the upper end of the
plug to see that the meniscus of the sample does not get very close
to this groove. Now spin the sample at high speed (typically position
3 on a clinical centrifuge) and the bubble will disappear. Now push
the upper plug down a tiny bit more and spin again and repeat this zero
to 2 more times. If you are especially worried about bubbles or have
an oxygen-sensitive sample, do the pushing in the gas bag under helium,
then put on the usual cap and spin outside. Finally spin several minutes
at high speed until preferably the sample meniscus is a few mm below
the groove on the plug. When this is achieved as well as seems possible,
store the sample spinning at the lowest speed until the epoxy is ready.
I use a 1/2 to 1 hour slow
setting household epoxy that comes in a double syringe, but throw away
the stupid storage plugs they supply. I use small dispo-pipette tips
instead, points shoved into the double-syringe holes. We store the double
epoxy syringe with its tip up on the flat end of the dual syringe, and
the syringe should be withdrawn after each use so that there is a small
bubble on each side before the pipette tips are put in. When mixing,
put a ml or so of epoxy in a small plastic weigh-boat. Be sure to put
the pipe tips back in the same side that they were when you removed
them, using their color or just keeping track or using new ones every
time. I generally have not degassed the epoxy before use but that could
be done. If bubbles may be a problem (for samples with small line-width)
I stir the epoxy in the gas bag under helium.
I use the #16 syringe instead
of the complicated procedure in MRC to transfer the epoxy to the top
of the upper plug. I forcefully push the plunger up while the end of
the tip is immersed in the newly mixed epoxy and get some epoxy to appear
as a blob above the top end of the syringe tip, inside the syringe body.
Then I apply it to the top of the upper plug, distributing it around
the circumference to trap gas in the groove space at the top of the
plug. Once this happens (normally very rapidly) I do not add much more
epoxy. I used to spin momentarily to bring down the epoxy on the walls,
usually by turning on the centrifuge at the lowest position for 15-30
sec only. If spun too long it might figure out how to sneak down into
the groove. This is no longer required with the new short tubes, because
you can get the end if the syringe right to the top of the plug. Finally,
inspect carefully to verify that the epoxy is above the groove and the
aqueous sample is below. If not, you could pull out the plug and throw
away the NMR tube and start over (but I seldom do. In the rare cases
where there seems to be a possibility of mixing, I usually just spin
(see above) and charge ahead, hoping for the best).
Now smear the remaining epoxy
on the upper 10 mm or so of the outside of the glass tube and the inside
of the adapter. You need enough to fill the space between the glass
and the adapter, but excess will have to be wiped off and also it could
get into the upper end of the hole and corrupt the thread there (this
has not yet happened). Shove them together holding the right side up
(tube to the bottom) until the center of the sample is 65 mm below the
lower end of the adapter (or whatever distance you find is needed).
Try to hold it there as you screw the adapter firmly (but only hand-tight)
into the shuttle body. Doing this seals the top fairly tightly and if
you get the sample in the wrong place, after screwing it on the body,
the trapped gas resists correcting the position and you have to unscrew
it partly to be able to move it to the desired position without force,
and then re-tighten.
Finally I place the assembled
parts in the vee-block, usually doing this with the vee block momentarily
horizontal to facilitate putting on the rubber bands and getting the
glass tube and body in the correct place approximately and not dropping
anything on the floor (usually). The lower end of the glass tube should
be even with the end of the vee block, and the 65 mm distance from sample
center to the adapter bottom should be set using a ruler. Then I tip
the V block on end and recheck these things and make sure that the above
adjustments are correct, and that the glass, the vee block adapter,
the nmr tube adapter, and the vee block are all pressed together by
the rubber bands. A folded-up piece of paper towel can be placed between
the rubber band and the NMR tube to push the tube in place more forcefully.
Check again later that all is well. If my lab is cold, in winter, I
do this in a warmer room. The assembly can be remove in about 3 hours
if you need to seal another tube, but when we shuttled once the same
day the epoxy failed. So we wait a day before using it. Also, after
removing it from the rubber bands, you can check by feeling with fingers,
or with a microscope (see MRC) to see if it fits in the vee-block +
vee block adapter without excess space and independent of the angular
position of the shuttle body. (for the 8 mm tube or a 5 mm tube to be
used in the 10 mm probe this is not so necessary because of the > 1
mm clearance. In this case I use an old-model spare shuttle to get good
enough alignment).
Bubbles can form in the sample
after running or after storage in the refrigerator. During running we
keep the down-travel faster than the up travel (down-pressure greater
than the up-suction) to inhibit bubble formation. This is not so serious
a restriction as you may think, because a delay in the initial part
of the up travel occurs where the sample is in high field and the T1
is longer. Usually we keep the vacuum half as large as the pressure
(both referenced to atmospheric pressure). We have to remember that
the vacuum gauge dial is marked in inches of Hg, and divide by two.
Breaking the NMR tubes
(?cleanly). I have a small rotary grinder (Dremel) mounted by a
small vise, that turns a small circular diamond saw, with a piece of
large aluminum I-beam that has a saw cut in it. These are mounted together
to resemble a minature circular wood saw. You cannot use this to entirely
cut the glass without a jagged result, but it is very useful for making
a series of usually 3 cuts that do not go through the glass. Then with
practice you can get a break that is only slightly jagged. These do
no great harm since the adapter is drilled out with excess length, and
the tube is pushed in to get the sample itself in the same place, as
described. As long as there is more than a cm of un-cracked tube at
the bottom end of the adapter to cement the tube in place, the jagtged
ends gives no problem.
Storage. Bubbles tend
to appear in the sample region during the night before we can run or
after the sample is stored in the refrigerator. I do not know what gas
they are or how they get there. I get rid of them by spinning the sample
in either our small clinical centrifuge or if that does not work, in
a large benchtop swinging bucket centrifuge of the type made by major
manufacturers. Whatever you do, it should be a swinging bucket. The
new short NMR tube and adapter will probably allow you to spin by just
putting the tube+adapter in a small-hole centrifuge tube carrier with
a counterbalance. If not, I have a thin-wall brass tube that fits into
the centrifuge carrier and around the tube with the adapter. It is long
enough so that when spinning it would swing down on the center part
of the rotor. But to start the centrifuge I have to close the top on
this brass tube. So I tie it down loosely with a cable tie (or any old
piece of wire, but the tie is better, and can be left there permanently,
un-noticed by the members of the owner's lab ). (I did get her permission
to do this however). The length of the brass tube has to be just right
for this to work.
Vee block and vee block
adapter. The vee block differs from the MRC version to make it easy
to align various diameters. It is now a single block of (surplus) aluminum
3" square by 14" long (see fig SH-2).
Great precautions were taken to mill it precisely and straight by clamping
it directly on the movable base of the milling machine and using a precision
45 degree cutter in good condition. It was outside-milled square first
and the outer edges were rounded to reduce breakage of the rubber bands.
A narrower rough square mill in the center (where the tip of V-mill
was going to be) allowed us to avoid having the tip of the V-mill do
any cutting that could impair its accuracy. Since the block is always
used with objects that are 0.8 inch in diameter it is only the part
that would touch this diameter that matters. I worry that the aluminum
will easily scratch. A brass or steel block, perhaps thinner and permanently
screwed to a thicker block or I-beam to avoid bending, might be worth
considering.
The vee block adapter
is a simple precision-turned brass block that is precisely the same
diameter as the glass tube inside (0.800 inch), and has a highly concentric
inside hole reamed no more than .001 inch larger than the 5 mm NMR tube
(fig. SH 2). It is then then cut in
half in a plane parallel to its axis and offset by about .01 inch from
it, and the slightly smaller half is carefully trimmed and used. (or
in another model we milled a 5 mm wide groove from the outer surface
toward the center so that a tube can be pushed in from the side. It
is roughly held to the vee block with tape for cementing the NMR tube
in place (rubber band are later used to hold it more firmly in place
after very thing is assembled), Test alignment by feeling with your
finger for play (to sense ~ 0.1 mm play) or use with a microscope (see
MRC). Different adapters can be easily made and used to align-mount
different round objects as precisely as possible. For an 8 mm NMR tube
that fits in a 10 mm probe alignment is not much of a problem and we
made the vee block adapter using the next larger available drill (US
21/64 inch) Note that an 8.00 mm reamer may be too small for an 8 mm
Wilmad tube whose minimum dimension is 8 mm.
Design and fabrication.
We do not use a separate lower susceptibility plug, use of epoxy (above)
seems satisfactory but we do not know its magnetic susceptibility. NMR
linewidths look ok for macromolecules. Contamination of the sample by
something from the epoxy (including oxygen) is certainly a concern but
we have never had any evidence of it. We have made solid PEEK lower
plugs for the 8 mm tubes but these little plugs have to be exoxy-cemented
into the tube, and are hard to clean for re-use.
You need as many upper plugs
(fig. SH 3 and fig.
SH 4) as the number of samples that you expect to keep stored at
any time, perhaps 4 to 10.
This section describes all
the parts of the system made of PEEK, which is very expensive. It is
used because it is very strong and hard. It is easy to clean hardened
epoxy from it, and the important interface surface, between it and the
precision glass tube that it slides inside of, never wears. It is not
very hard to machine, but all work has to be done with carbide-tipped
steel.
The upper plug (fig.
SH-3, fig. SH-4) was redesigned
with moderately tight clearance (.06 mm nominal) for the body (for the
central tube, Fig. SH-4), and with
end caps turned to as close a fit as possible to still allow the plug
to be pushed freely into the tube. Four flats were index-milled on the
lower cap to allow the sample to pass when the plug is pushed in, to
the same diameter (Fig. SH-3). Our
theory (see MRC) (which is not well verified) is that the ends keep
the central tube aligned so that sloshing, produced by the >1000 G deceleration
at the end of each travel, and partly in cooperation with un-evenness
in the thickness of the plastic-to-glass clearance, will be inhibited.
Originally I filed the flats by hand, using a magic-marker as well as
a micrometer to keep track of where to file.
The upper plug also carries
the groove at its top that is shallower than that described in MRC (now
0.13 mm deep, 1 mm long). We went back to using this groove because
we were worried about water-epoxy mixing problems. The smaller-volume
groove reduces the compressibility that the fluid sample sees. This
compressibility may help the sloshing to occur. We may try an even smaller
groove; the current groove nearly always keeps the setting epoxy away
from the solution.
The central tube of the upper
plug (Fig SH-4) is made from 1/4" PEEK
stock. We bought a 3-fluted carbide steel drill to make the hole, and
the hole is drilled half-way from each end with a thick section in the
middle to provide extra stiffness. Great care is needed to avoid variations
in the outside diameter of the plug, that might come from machining
too fast.
The caps are glued with household
1/2 hour epoxy. It is desirable to machine the joints between the caps
and the central tube to fit well, but not so tightly that they expand
the outer diameter of the central tube. The wire at the top is usually
epoxied at the same time. Do not be alarmed if the wire and/or the caps
come apart during re-cycling of the plug when it is broken out of the
glass tube and cleaned of old epoxy, simply re-epoxy them (but clean
the cap a day in advance to allow time for this). Excess epoxy is wiped
of with a tissue (Kimwipe) before it is set, and scraped off with a
razor blade afterward (see above) and tested in an a NMR tube to see
that it slide easily. The wire is bent and tested to give moderate resistance
to sliding exery time it is used.
The upper plug is now (as
of 5/2004) rather long (~35 mm), in order that the assembled plug would
float during spinning, before we used the wire at the top as a spring
to keep the plug in place. A length shorter than this may be desirable
and the floatation feature may not be needed, we do not know at this
point and we will try shorter plugs (1.5 cm?). The reason why we want
to try this (aside from the fact that the plug will be easier to make)
is that when assembled to the adapter (Fig.
SH1) the upper end of the plug must be a few millimeters or more
below the end of the adapter. Otherwise it is hard to disassemble the
plug from the adapter without destroying it. If the nmr tube were shorter
by about a cm, then the distance from the sample center to the adapter
bottom could be reduced to about 55 mm instead of the present 65 mm
and the tube will be shorter and stiffer. The distance of 55 mm was
our original design, and would allow the bottom of the adapter to be
about 5 mm above the recessed plastic top surface of the probe. The
bottom of the adapter is now even with the bottom of the shuttle tube
holder assembly (below) when it is at the bottom of its travel (Fig.
ST2).
Other variations of the plug
design will be tried, perhaps a nearly solid short design. It is useful
but not essential to be able to be able to reach in and pull out the
upper plug in case of some error as we now can, but it is not essential.
We even connected a piece of thread to the wire at one point to be able
to do this but gave it up since we never used it.
You need as many NMR tube
adapters (Fig SH5, not to be confused
with vee-block adapters above) as you do upper plugs. They are designed
to be made easily and are made from 1/2" PEEK stock turned down to slightly
smaller diameter so that the upper ends can be machined to get good
and defined alignment to mate with the lower end of the shuttle body.
Note that the upper central shoulder (0.3 inch diameter, at the top
of the adapter) is less high than the depth of the corresponding cylindrical
hole that it mates with(see Fig SH-7),
so that it does not touch this surface of the shuttle body when it and
the adapter are screwed together, and only the outer annular surfaces
of the adapter and the body ends are used for alignment to perpendicularity,
and the diameters around the inside of these surfaces are used for radial
alignment. The mate should be rather snug, and the fit should be such
that there is no play before screwing together, and no big crack visible
where the two pieces come together, afterward. The edge of the connection
surface marked (Fig. SH-5) should be
rounded slightly to allow this. Although the assembly of the NMR tube
using the vee block etc. is the most important aspect of our system
to get good alignment, we also think it desirable that the adapters
and their connection to the shuttle body should be excellent.
The stud connecting the adapter
to the shuttle body is made without special care from a cut-off brass
or stainless steel screw and trimmed. It is designed to be screwed moderately
tightly into the shuttle body, where it will live forever. If it is
not brass, check it with a stir-bar or other magnet for non-attraction.
The long hollow tube of the
shuttle body (Fig. SH-6) does not require
great care in fabrication for alignment. The design shows an optional
thick section in the middle if you wish to drill from each end (we did
not do this the first time). It is now shorter than that in MRC but
could be longer than shown. We bought a carbide-tipped drill of an inexpensive
type (Koolcar durapoint, MSC catalog p.95) to drill this hole from 1"
stock, after finishing the outside and predrilling with a ½" carbide
tipped drill. Then we turned a 1/2 " long section in the inside diameter
at each end of the tube for the epoxy joint with the two end pieces,
slightly larger than the drill hole. It is a good idea to score these
surfaces and the mating ones lightly after this, to make the epoxy permanent
Then we turned the outside diameter to ~3/4 inch, to reduce the weight.
The two end caps (fig. SH7)
carry the 0.798 nominal dia. bearing surfaces that have .001" (.03 mm)
nominal clearance with the precision glass shuttle tube. This precision
surface should be made, for the bottom cap, at the same time as the
lower end of the lower cap that mates with the NMR adapter, to make
them as well aligned as is easily possible. The 1/2 " outer surfaces
that carry the seven O-rings are not so critical. I worried that the
seven O-rings would fall off but they do not seem to, they are actually
slightly less than ½ " inside diameter. Be sure to get Viton (Fluorocarbon)
O rings for this.
You can cement (with 1/2
hour epoxy) the lower cap while the tube is in a collet in the lathe,
setting up an indicator to test for alignment. Then after it has set,
cement the other end using a jaw holder in the same way. It is probably
better to epoxy the top cap first, without special precautions, wait
for it to set, then epoxy the bottom cap using the vee block. First
make a vee-block adapter with the correct diameter (slightly less than
1/2") to fit the outside of the NMR adapters. Then screw the adapter
and bottom cap together before epoxying the bottom cap. Do this epoxy-ing
in the same way as we seal the NMR tube, with the vee-block.
We test the vee-block procedure
as described in MRC.
The shuttle should drop freely
though the open glass tube in about 1 sec but, with an adapter screwed
on, it falls remakably slowly, abou1 cm/sec.
Temperature. See MRC.
We have only done a few longer runs up to around 50° C. I expect from
various experiences that cracking of the epoxy, or the glass at the
epoxied points, will occur at some temperature not much above this.
I think it is better to use as little epoxy as possible. Possibly the
nmr tube seals could be set in a warm room or incubator at the mid-point
of the desired temperature range. Possibly a supplier like Epoxy Technology
could suggest a good epoxy. (However, the high viscosity of the household
epoxy we now use is good for cementing the upper plug in the glass tube.
Don't substitute a lower-viscosity epoxy here.)